4.

OPERATION

Practically all procedures for measuring noise dose require that the
measuring device be calibrated both before and after the measurement !

It is also recommended that the current configuration be checked

(see Section 2.5) to ensure that it is suitable for the proposed
measurement task. If the configuration is not suitable, change it before
commencing calibration or operation, as described in Chapter 3.

4.1

Automatic Calibration

With Dose Badge internal batteries charged as described in Section 2.3
and batteries installed in the Reader Unit
(Section 2.4), calibrate all Dose Badges to be
used for the proposed measurement task as
follows.

1.

Carefully insert the Dose Badge into
the calibrator cavity in the top panel
of the Reader, as shown in Figure 5.

Make sure the Dose Badge is firmly
in contact with the shoulder near the
bottom of the cavity.

Also make sure the infrared window
in the Dose Badge is facing the
window in the top panel of the
Reader so that effective
communication will be possible.

2.

Press

to switch the Reader ON.

3.

Wait for the status screen to be

displayed, then press

to initiate

an automatic calibration sequence.

Cal

C-01

Calibrating . . . .

Battery condition

Calibration progress

03022

Figure 5: Dose Badge

installed in the calibration

cavity with infrared

windows aligned for

communication

Operation

Page 20 - CEL-310 Dose Badge Users Handbook