MicroDust Pro Aerosol Monitoring System

10

Casella CEL

Signals

A 6-way port at the bottom of the MicroDust Pro instrument provides the
following connections;

RS232 serial digital communications for connecting the instrument to a

PC

An analogue output signal for connecting the instrument to a chart

recorder or to an external data logger.

Table 1.

6-way port

Pin

Wire

colour

9-way D-type

pin and function

Function

1

Yellow

N/A

Analogue ground

2

Red

N/A

Analogue out

3

White

3 (TX out)

RS232 receive in

4

Black

2 (Rx in)

RS232 transmit

out

5

Blue

5 (Ground)

Comms ground

4 (DTR

8 (CTS)

6 (DSR)

Linked together

The analogue output signal has a full-scale range of 0 to 2.5 volts and an output
impedance of approximately 500 Ω. The full-scale output voltage is
proportional to the full-scale indication of the active range. For example, if the
active range is 0 to 25 mg/m³, and output signal of 2.5 volts represents a dust
concentration of 25 mg/m³ (meaning the analogue output is scaled at 1 mV per
0.01 mg/m³).

IMPORTANT

The analogue output is available only when using a fixed measurement range (that is,

when NOT using Auto Range).