MicroDust Pro Aerosol Monitoring System

38

Casella CEL

TSP and Respirable adapters

Figure 20.

Gravimetric and respirable adapter modules

1 Pump connection

8 Measuring chamber inlet

2 Adapter block

9 Locating spigot

3 Locating spigot

10 Cyclone

4 Inlet tube

11 Coarse dust container

5 Thumb screw

12 Filter cassette (25 mm / 37 mm)

6 37 mm-to-25 mm reducer

13 O-rings

7 MicroDust sampling probe

The basic design of the TSP and respirable adapters is the same, but with
different inlet fittings. The respirable adapter is a centrifugal separator that
uses the cyclone principle of dust separation. It draws only the respirable
fraction of particulate matter through the measurement chamber and onto the
filter (D50 cut = 4 µm at 2.2 litres/minute).