MicroDust Pro Aerosol Monitoring System

5

Casella CEL

Figure 1.

Sampling probe

1.

Infra-red emitter

2.

Sample volume

3.

Infra-red detector

4.

Light stop

Under clean-air conditions, all light is prevented from reaching the received by
a light stop (4). When dust particles enter the sample volume (2), the beam of
infra-red light is scattered within a narrow angle (12° to 20°) to the receiver (4).

By using a narrow angle of scatter, the instrument is made less sensitive to
variations in the refractive index and colour of measured particulate.