Casella CEL Microdust Pro real time dust monitor User Manual

Page 20

Advertising
background image

MicroDust Pro Aerosol Monitoring System

20

Casella CEL

Figure 10.

Setting the Averaging Time

The displayed particulate concentration is a rolling average of measurements
taken over the averaging time that you set. In some applications, where aerosol
concentrations change rapidly, you can improve measurement stability by
setting a longer averaging time.

Analogue output type

The analogue output type is optional. Choose from the following options.

Raw

The analogue output is taken directly from the signal
conditioning circuit and is normalised to match the
instrument’s measurement range.

Average

The analogue output is calculated using the averaging time you
have set for the instrument.

Alarm

The output voltage represents a set digital value where 2.5 V is
Alarm ON and 0 V is Alarm OFF.

Alarm

If you set the Alarm option to ON, you can set the following options.

Alarm Level

Set the alarm trigger level as xxx.xx mg/m³.

Press and hold the  or  key to change the setting quickly.