MicroDust Pro Aerosol Monitoring System

42

Casella CEL

The aspirator fan operates from a 5 V DC power supply, which can come from a
suitable mains DC adapter or from a battery supply.

You should clean the aspirator regularly to make sure there is no accumulation
of dust, which can cause measurements errors or mechanical seizure of the fan.

7

Specifications

7.1

Instrument specification

Measurement

Sensing technique

Forward light scattering (12° to 20°) using infrared
source (wavelength 880 nm)

Measuring range

0.001 to 2500 mg/m³ over four ranges:

0 to 2.5 mg/m³

0 to 25 mg/m³

0 to 250 mg/m³

0 to 2500 mg/m³

The active range can be fixed or auto ranging.

Resolution

0.001 mg/m³

Zero stability

Typically <2 µg/m³/°C

Span stability

±0.7 % of reading/°C

Temperature:

Operating 0 °C to 55 °C (non-condensing)

Storage –20 °C to +55 °C (non-condensing)

Power supply

Power consumption

Approximately 100 mA at 6 V

Battery

Four AA / MN1500 cells
Alkaline or rechargeable NiMH

Operating run time on batteries:

Alkaline (2700 mAh) Typically >20 hours

NiMH (950 mAh) Typically >9 hours