Using the instrument

The MicroDust Pro is a sensitive scientific instrument that must be protected
against excessive moisture, vibration and physical shock. Follow the
instructions in this User Manual carefully to avoid causing damage to the
instrument.

This section of the manual describes the tasks you must perform to make the
MicroDust Pro instrument ready to make measurements. You should try to
complete the tasks in the sequence they are included.

Preparing the instrument for use

Preliminary operations

Calibration

IMPORTANT

For optimum accuracy, you should always perform a user calibration to correct your

instrument for the type of particulate to be measured. Refer to

Calibrating the

instrument for a specific type of particulate

on page 25 for instructions to do this.

Making measurements

Data logging

3.1

Preparing the instrument for use

Setting the DIP switches

WARNING
DO NOT enable the charger circuit while there are non-rechargeable batteries installed

in the MicroDust Pro instrument.

The MicroDust Pro instrument has a built-in fast-charging circuit that uses
power supplied by the 12 V DC supply. This allows you to recharge suitable
rechargeable batteries without the need to remove them from the instrument.

Before you can use this feature, you must set the Charger Enable DIP switch
to the ON position, as shown by item (2) in

Figure 4

. The DIP switches are

located in the battery compartment.

If you use non-rechargeable batteries to operate the instrument, it is very
important to set the Charger Enable DIP switch to the OFF position.