Calibration
Elgar Electronics Corporation
2.3 VSENSE
CALIBRATION
Make the connections for VSENSE calibration as shown in Figure 2-2.
Figure 2-2. Connections for Vsense Calibration (Items 5-8).
2.3.1
VS
ENSE
, R
ANGE
I O
FFSET AND
G
AIN
, I
TEMS
5
AND
6
1.
Press the
© key to go to Item 5.
2.
Set the DC Voltage Standard to the Input Voltage value shown in
Table 2-5 for the model being calibrated.
3.
Use the
keys to adjust the load DVM display until it matches the
DMM reading.
MODEL
Input Voltage
MODEL
Input Voltage
SLM-60-30-150 0.200V SLH-60-120-600 0.200V
SLM-60-60-300 0.200V SLH-60-120-1200
0.200V
SLM-250-10-300 0.200V SLH-60-120-1800
0.200V
SLM-500-10-300 0.400V SLH-60-240-1200
0.200V
SLM-60-15-75 0.200V SLH-60-240-1800
0.200V
SLH-60-360-1800
0.200V
SLH-500-60-1800
0.400V
Table 2-5. VSense, Range I Offset Adjustment Settings (Item 5)
4.
Press the
© key to go to Item 6.
2-4
SLM/SLH Calibration Manual