2.3 VSENSE

CALIBRATION

Make the connections for VSENSE calibration as shown in Figure 2-2.

Figure 2-2. Connections for Vsense Calibration (Items 5-8).

2.3.1

VS

ENSE

, R

ANGE

I O

FFSET AND

G

AIN

, I

TEMS

5

AND

6

1.

Press the

© key to go to Item 5.

2.

Set the DC Voltage Standard to the Input Voltage value shown in
Table 2-5 for the model being calibrated.

3.

Use the

…† keys to adjust the load DVM display until it matches the

DMM reading.

MODEL

Input Voltage

MODEL

Input Voltage

SLM-60-30-150 0.200V SLH-60-120-600 0.200V
SLM-60-60-300 0.200V SLH-60-120-1200

0.200V

SLM-250-10-300 0.200V SLH-60-120-1800

0.200V

SLM-500-10-300 0.400V SLH-60-240-1200

0.200V

SLM-60-15-75 0.200V SLH-60-240-1800

0.200V

SLH-60-360-1800

0.200V

SLH-500-60-1800

0.400V

Table 2-5. VSense, Range I Offset Adjustment Settings (Item 5)

4.

Press the

© key to go to Item 6.

2-4