Elgar Electronics Corporation

Calibration

9.

Input Voltage to the VSENSE terminal per Table 2-34 for the model
being calibrated.

10.

PRES ON, setting Power as table 36 for the model being calibrated.

11.

PRES OFF.

12.

LOAD ON, Adjust using the

…† keys until the load DAM matches

Current Setting value.

13.

LOAD OFF.

14.

Repeat steps 1 through 13 (press

ª to return to Item 21) until no

further adjustments are necessary for the load DAM to match the
DMM reading.

MODEL

Power Setting

VSENSE input

Current Setting

SLM-60-30-150 150.0W 50.00V 3.000A
SLM-60-60-300 300.0W 50.00V 6.000A
SLM-250-10-300 300.0W 200V 1.500A
SLM-500-10-300 300.0W 300V 1.000A
SLM-60-15-75 75.0W

50.00V

1.5000A

SLH-60-120-600 600.0W 50.00V 12.00A
SLH-60-120-1200 1200.0W 50.00V 24.00A
SLH-60-120-1800 1800.0W 50.00V 36.00A
SLH-60-240-1200 1200.0W 50.00V 24.00A
SLH-60-240-1800 1800.0W 50.00V 36.00A
SLH-60-360-1800 1800.0W 50.00V 36.00A
SLH-500-60-1800 1800.0W 300.00V 6.000A

Table 2-34. CP Mode, Level Low, Gain Adjustment Settings (Item 34)

SLM/SLH Calibration Manual

2-31