AMETEK SL Series Calibration Manual User Manual
Page 19
Elgar Electronics Corporation
Calibration
8.
LOAD OFF.
9.
Press the
© key to go to Item 14.
10.
PRES ON.
11.
Set the load current as shown in Table 2-16 for the model being
calibrated.
12.
PRES OFF.
13.
LOAD ON.
14.
Adjust using the
keys until the shunt current and the current
setting match.
15.
Repeat steps 1 through 14 (press
ª to return to Item 13) until no
further adjustments are necessary for the shunt current and the
current setting to match.
MODEL
Current Setting
MODEL
Current Setting
SLM-60-30-150 3.000A SLH-60-120-600 12.00A
SLM-60-60-300 6.000A SLH-60-120-1200
12.00A
SLM-250-10-300 1.0050A SLH-60-120-1800
12.00A
SLM-500-10-300 1.0050A SLH-60-240-1200
24.00A
SLM-60-15-75 1.500A SLH-60-240-1800
24.00A
SLH-60-360-1800
36.00A
SLH-500-60-1800
6.000A
Table 2-14. CC Mode, Level Low, Range I Gain Adjustment Settings (Item 14)
SLM/SLH Calibration Manual
2-11