Calibration

Elgar Electronics Corporation

2.4.4

CC M

ODE

, L

EVEL

H

IGH

, R

ANGE

I O

FFSET AND

G

AIN

, I

TEMS

15

AND

16

1.

LOAD OFF.

2.

Press the

© key to go to Item 15.

3.

PRES ON.

4.

Set the load current as shown in Table 2-15 for the model being
calibrated.

5.

PRES OFF.

6.

LOAD ON.

7.

Adjust using the

…† keys until the shunt current and the current

setting match.

MODEL

Current

Setting MODEL Current

Setting

SLM-60-30-150 0.00A

SLH-60-120-600 0.00A

SLM-60-60-300 0.00A

SLH-60-120-1200 0.00A

SLM-250-10-300 0.00A

SLH-60-120-1800 0.00A

SLM-500-10-300 0.00A

SLH-60-240-1200 0.00A

SLM-60-15-75 0.00A

SLH-60-240-1800

0.00A

SLH-60-360-1800

0.00A

SLH-500-60-1800

0.00A

Table 2-15. CC Mode, Level High, Range I Offset Adjustment Settings (Item 15)

8.

LOAD OFF.

9.

Press the

© key to go to Item 16.

10.

PRES ON.

11.

Set the load current as shown in Table 2-16 for the model being
calibrated.

12.

PRES OFF.

13.

LOAD ON.

14.

Adjust using the

…† keys until the shunt current and the current

setting match.

15.

Repeat steps 1 through 14 (press

ª to return to Item 15) until no

further adjustments are necessary for the shunt current and the
current setting to match.

2-12

SLM/SLH Calibration Manual