7 cp mode calibration, Cp m, Alibration – AMETEK SL Series Calibration Manual User Manual

Page 37

Advertising
background image

Elgar Electronics Corporation

Calibration

10.

PRES ON.

11.

Set voltage to the V Input value shown in Table 2-32 for the model
being calibrated.

12.

PRES OFF.

13.

LOAD ON.

14.

Adjust using the

…† keys until the DVM reading and the Voltage

Setting match.

15.

Repeat steps 1 through 14 (press

ª to return to Item 30) until no

further adjustments are necessary for the DVM reading and the
Voltage Setting to match.

MODEL

V Input Voltage Setting

MODEL

V Input Voltage Setting

SLM-60-30-150 61.5 60.000V SLH-60-120-600 61.5 60.000V
SLM-60-60-300 61.5 60.000V SLH-60-120-1200

61.5 60.000V

SLM-250-10-300 252

250.0V SLH-60-120-1800

61.5

60.000V

SLM-500-10-300 505

500.2V SLH-60-240-1200

61.5

60.000V

SLM-60-15-75 61.5 60.000V SLH-60-240-1800

61.5 60.000V

SLH-60-360-1800

61.5

60.000V

SLH-500-60-1800

505

500.2V

Table 2-32. CV Mode, Level High, Gain Adjustment Settings (Item 32)

2.7 CP

MODE

CALIBRATION

Make the connections shown in Figure 2-6.

Figure 2-6. Connections for CP Mode Calibration

SLM/SLH Calibration Manual

2-29