AMETEK SL Series Calibration Manual User Manual

Page 18

Advertising
background image

Calibration

Elgar Electronics Corporation

10.

PRES ON.

11.

Set the load current as shown in Table 2-12 for the model being
calibrated.

12.

PRES OFF.

13.

LOAD ON.

14.

Adjust using the

…† keys until the shunt current and the load DAM

reading match.

15.

Repeat steps 1 through 14 (press

ª to return to Item 11) until no

further adjustments are necessary for the load DAM to match the
DMM reading.

MODEL

Current Setting

MODEL

Current Setting

SLM-60-30-150 30.00A

SLH-60-120-600 120.00A

SLM-60-60-300 60.00A

SLH-60-120-1200

120.00A

SLM-250-10-300 10.050A

SLH-60-120-1800 120.00A

SLM-500-10-300 10.050A

SLH-60-240-1200 240.00A

SLM-60-15-75 15.000A

SLH-60-240-1800

240.00A

SLH-60-360-1800

360.00A

SLH-500-60-1800

60.00A

Table 2-12. DAM, Range II Gain Adjustment Settings (Item 12)

2.4.3

CC M

ODE

, L

EVEL

L

OW

, R

ANGE

I O

FFSET AND

G

AIN

, I

TEMS

13

AND

14

1.

LOAD OFF (Power supply voltage should still be at +5Vdc).

2.

Press the

© key to go to Item 13.

3.

PRES ON.

4.

Set the load current as shown in Table 2-13 for the model being
calibrated.

5.

PRES OFF.

6.

LOAD ON.

7.

Adjust using the

…† keys until the shunt current and the current

setting match.

MODEL

Current Setting

MODEL

Current Setting

SLM-60-30-150 0.00A

SLH-60-120-600 0.00A

SLM-60-60-300 0.00A

SLH-60-120-1200 0.00A

SLM-250-10-300 0.00A

SLH-60-120-1800 0.00A

SLM-500-10-300 0.00A

SLH-60-240-1200 0.00A

SLM-60-15-75 0.00A

SLH-60-240-1800

0.00A

SLH-60-360-1800

0.00A

SLH-500-60-1800

0.00A

Table 2-13. CC Mode, Level Low, Range I Offset Adjustment Settings (Item 13)

2-10

SLM/SLH Calibration Manual

Advertising