AMETEK SL Series Calibration Manual User Manual

Page 21

Advertising
background image

Elgar Electronics Corporation

Calibration

16.

PRES ON.

17.

Set the load current as shown in Table 2-16 for the model being
calibrated.

18.

PRES OFF.

19.

LOAD ON.

20.

Adjust using the

…† keys until the shunt current and the current

setting match.

21.

Repeat steps 1 through 14 (press

ª to return to Item 15) until no

further adjustments are necessary for the shunt current and the
current setting to match.

Current

Setting

MODEL

MODEL Current

Setting

SLM-60-30-150 3.000A

SLH-60-120-600 12.00A

SLM-60-60-300 6.000A

SLH-60-120-1200

12.00A

SLM-250-10-300 1.0050A

SLH-60-120-1800

12.00A

SLM-500-10-300 1.0050A

SLH-60-240-1200

24.00A

SLM-60-15-75 1.500A

SLH-60-240-1800

24.00A

SLH-60-360-1800

36.00A

SLH-500-60-1800

6.000A

Table 2-16. CC Mode, Level High, Range I Gain Adjustment Settings (Item 16)

2.4.5

CC M

ODE

, L

EVEL

L

OW

, R

ANGE

II O

FFSET AND

G

AIN

, I

TEMS

17

AND

18

1.

LOAD OFF.

2.

Press the

© key to go to Item 17.

3.

PRES ON.

4.

Set the load current as shown in Table 2-17 for the model being
calibrated.

5.

PRES OFF.

6.

LOAD ON.

7.

Adjust using the

…† keys until the shunt current and the current

setting match.

MODEL

Current Setting

MODEL

Current Setting

SLM-60-30-150 0.00A

SLH-60-120-600 0.00A

SLM-60-60-300 0.00A

SLH-60-120-1200 0.00A

SLM-250-10-300 0.00A

SLH-60-120-1800 0.00A

SLM-500-10-300 0.00A

SLH-60-240-1200 0.00A

SLM-60-15-75 0.00A

SLH-60-360-1800

0.00A

SLH-500-60-1800

0.00A

Table 2-17. CC Mode, Level Low, Range II Offset Adjustment Settings (Item 17)

SLM/SLH Calibration Manual

2-13