Calibration

Elgar Electronics Corporation

11.

Set the power supply voltage and the DC Voltage Standard to the
VSENSE Input value shown in Table 2-22 for the model being
calibrated.

12.

PRES ON.

13.

Set the current to the Current Setting value in Table 2-22 for the
model being calibrated.

14.

PRES OFF.

15.

LOAD ON.

16.

Adjust using the

…† keys until the DAM reading and the current

setting match.

17.

Repeat steps 1 through 16 (press

ª to return to Item 21) until no

further adjustments are necessary for the DAM reading t and the
current setting to match.

MODEL Resistance

Setting

VSENSE Input

Current Setting

SLM-60-30-150 2.000

Ω 60.00V 30.00A

SLM-60-60-300 1.000Ω 60.00V 60.00A
SLM-250-10-300 25.00Ω 250.00V 10.00A
SLM-500-10-300 50.00Ω 500.00V 10.000A
SLM-60-15-75 4.000Ω 60.00V 15.000A

SLH-60-120-600 0.5000

Ω 60.00V 120.00A

SLH-60-120-1200 0.5000

Ω 60.00V 120.00A

SLH-60-120-1800 0.5000

Ω 60.00V 120.00A

SLH-60-240-1200 0.2500

Ω 60.00V 240.00A

SLH-60-240-1800 0.2500

Ω 60.00V 240.00A

SLH-60-360-1800 0.167

Ω 60.00V 359.281A

SLH-500-60-1800 8.3333

Ω 500.00V 60.00A

Table 2-22. CR Mode, Level Low, Range I Gain Adjustment Settings (Item 22)

2-18