Calibration

Elgar Electronics Corporation

11.

Set the power supply voltage and the DC Voltage Standard to the
VSENSE Input value shown in Table 2-28 for the model being
calibrated.

12.

PRES ON.

13.

Set the current to the Current Setting value in Table 2-28 for the
model being calibrated.

14.

PRES OFF.

15.

LOAD ON.

16.

Adjust using the

…† keys until the DAM reading and the current

setting match.

17.

Repeat steps 1 through 16 (press

ª to return to Item 27) until no

further adjustments are necessary for the DAM reading and the
current setting to match.

MODEL

Resistance Setting

VSENSE

Current Setting

SLM-60-30-150 2.000Ω 60.00V

30.000A

SLM-60-60-300 1.0001Ω 60.00V

60.000A

SLM-250-10-300 25.00Ω 250.00V

10.000A

SLM-500-10-300 50.000Ω 500.00V

10.000A

SLM-60-15-75 4.000Ω 60.00V

15.000A

SLH-60-120-600 0.5000

Ω 60.00V

120.00A

SLH-60-120-1200 0.5000

Ω 60.00V

120.00A

SLH-60-120-1800 0.5000

Ω 60.00V

120.00A

SLH-60-240-1200 0.2500

Ω 60.00V

240.00A

SLH-60-240-1800 0.2500

Ω 60.00V

240.00A

SLH-60-360-1800 0.167

Ω 60.00V

359.281A

SLH-500-60-1800 8.3333

Ω 500.00V 60.00A

Table 2-28. CR Mode, Level High, Range II Gain Adjustment Settings (Item 28)

2-24

SLM/SLH Calibration Manual