8 store calibration data, Tore, Alibration – AMETEK SL Series Calibration Manual User Manual

Page 41

Advertising
background image

Elgar Electronics Corporation

Calibration

9.

Input Voltage to the VSENSE terminal as Table 2-36.

10.

PRES ON, setting Power as table 40, PRES OFF.

11.

LOAD ON, Adjust using the

…† keys until load DAM matches

Current Setting value.

12.

Repeat steps 1 through 11 (press

ª to return to Item 21) until no

further adjustments are necessary for the load DAM to match the
DMM reading.

13.

LOAD OFF.

MODEL

Power Setting

VSENSE Input

Current Setting

SLM-60-30-150 150W

5.00V

30.000A

SLM-60-60-300 300W 5.00V 60.000A
SLM-250-10-300 300W 30.00V 10.000A
SLM-500-10-300 300W 30.00V 10.000A
SLM-60-15-75 75W 5.00V

15.000A

SLH-60-120-600 600.0W 5.000V 120.0A
SLH-60-120-1200 1200.0W 10.000V 120.0A
SLH-60-120-1800 1800.0W 15.000V 120.0A
SLH-60-240-1200 1200.0W 5.000V

240.0A

SLH-60-240-1800 1800.0W 7.500V

240.0A

SLH-60-360-1800 1800.0W 5.000V

360.0A

SLH-500-60-1800 1800.0W 30.00V

60.00A

Table 2-36. CP Mode, Level High, Gain Adjustment Settings (Item 36)

2.8

STORE CALIBRATION DATA

Press SHORT key to store the calibration data.

Press LIMIT key to exit the calibration mode.

SLM/SLH Calibration Manual

2-33