AMETEK SL Series Calibration Manual User Manual

Page 25

Advertising
background image

Elgar Electronics Corporation

Calibration

2.5.1

CR M

ODE

, L

EVEL

L

OW

, R

ANGE

I O

FFSET AND

G

AIN

, I

TEMS

21

AND

22

1.

LOAD OFF.

2.

Press the

© key to go to Item 21.

3.

Set the power supply voltage and the DC Voltage Standard to the
VSENSE Input value shown in Table 2-21 for the model being
calibrated.

4.

PRES ON

5.

Set the current to the Current Setting value in Table 2-21 for the
model being calibrated.

6.

PRES OFF.

7.

LOAD ON.

8.

Adjust using the

…† keys until the DAM reading and the current

setting match.

MODEL Resistance

Setting

VSENSE Input

Current Setting

SLM-60-30-150 3750

Ω 60.00V 0.016A

SLM-60-60-300 3750

Ω 60.00V 0.016A

SLM-250-10-300 15625

Ω 250.00V 0.016A

SLM-500-10-300 6250

Ω 500.0V 0.08A

SLM-60-15-75 5000

Ω 60.00V 0.012A

SLH-60-120-600 1875

Ω 60.00V 0.0320A

SLH-60-120-1200 1875

Ω 60.00V 0.0320A

SLH-60-120-1800 1875

Ω 60.00V 0.0320A

SLH-60-240-1200 937.50

Ω 60.00V 0.0640A

SLH-60-240-1800 937.50

Ω 60.00V 0.0640A

SLH-60-360-1800 625

Ω 60.00V 0.0960A

SLH-500-60-1800 15625.

Ω 500.00V 0.0320A

Table 2-21. CR Mode, Level Low, Range I Offset Adjustment Settings (Item 21)

9.

LOAD OFF.

10.

Press the

© key to go to Item 22.

SLM/SLH Calibration Manual

2-17