AMETEK SL Series Calibration Manual User Manual
Page 22
Calibration
Elgar Electronics Corporation
8.
LOAD OFF.
9.
Press the
© key to go to Item 18.
10.
PRES ON.
11.
Set the load current as shown in Table 2-18 for the model being
calibrated.
12.
PRES OFF.
13.
LOAD ON.
14.
Adjust using the
keys until the shunt current and the current
setting match.
15.
Repeat steps 1 through 14 (press
ª to return to Item 17) until no
further adjustments are necessary for the shunt and the current
setting to match.
MODEL
Current Setting
MODEL
Current Setting
SLM-60-30-150 30.00A
SLH-60-120-600
120.00A
SLM-60-60-300 60.00A
SLH-60-120-1200
120.00A
SLM-250-10-300 10.050A
SLH-60-120-1800
120.00A
SLM-500-10-300 10.050A
SLH-60-240-1200
240.00A
SLM-60-15-75 15.000A
SLH-60-240-1800
240.00A
SLH-60-360-1800 360.00A
SLH-500-60-1800 60.00A
Table 2-18. CC Mode, Level Low, Range II Gain Adjustment Settings (Item 18)
2-14
SLM/SLH Calibration Manual