Calibration

Elgar Electronics Corporation

10.

PRES ON.

11.

Set the load current as shown in Table 2-20 for the model being
calibrated.

12.

PRES OFF.

13.

LOAD ON.

14.

Adjust using the

…† keys until the shunt current and the current

setting match.

15.

Repeat steps 1 through 14 (press

ª to return to Item 19) until no

further adjustments are necessary for the shunt current and the
current setting to match.

MODEL

Current Setting

MODEL

Current Setting

SLM-60-30-150 30.00A SLH-60-120-600 120.00A
SLM-60-60-300 60.00A SLH-60-120-1200 120.00A
SLM-250-10-300 10.050A SLH-60-120-1800

120.00A

SLM-500-10-300 10.050A SLH-60-240-1200

240.00A

SLM-60-15-75 15.000A SLH-60-240-1800 240.00A

SLH-60-360-1800 360.00A

SLH-500-60-1800 60.00A

Table 2-20. CC Mode, Level High, Range II Gain Adjustment Settings (Item 20)

2.5

CR MODE CALIBRATION

Connect the DC Power supply and DC Voltage Standard, as shown in
Figure 2-4.

Figure 2-4. Connections for CR Mode Calibration

2-16

SLM/SLH Calibration Manual