5 cr mode calibration, Cr m, Alibration – AMETEK SL Series Calibration Manual User Manual
Page 24
Calibration
Elgar Electronics Corporation
10.
PRES ON.
11.
Set the load current as shown in Table 2-20 for the model being
calibrated.
12.
PRES OFF.
13.
LOAD ON.
14.
Adjust using the
keys until the shunt current and the current
setting match.
15.
Repeat steps 1 through 14 (press
ª to return to Item 19) until no
further adjustments are necessary for the shunt current and the
current setting to match.
MODEL
Current Setting
MODEL
Current Setting
SLM-60-30-150 30.00A SLH-60-120-600 120.00A
SLM-60-60-300 60.00A SLH-60-120-1200 120.00A
SLM-250-10-300 10.050A SLH-60-120-1800
120.00A
SLM-500-10-300 10.050A SLH-60-240-1200
240.00A
SLM-60-15-75 15.000A SLH-60-240-1800 240.00A
SLH-60-360-1800 360.00A
SLH-500-60-1800 60.00A
Table 2-20. CC Mode, Level High, Range II Gain Adjustment Settings (Item 20)
2.5
CR MODE CALIBRATION
Connect the DC Power supply and DC Voltage Standard, as shown in
Figure 2-4.
Figure 2-4. Connections for CR Mode Calibration
2-16
SLM/SLH Calibration Manual