Elgar Electronics Corporation

2.7.2

CP M

ODE

, L

EVEL

H

IGH

, O

FFSET AND

G

AIN

, I

TEMS

35

AND

36

1.

LOAD OFF

2.

Press the

© key to go to Item 35.

3.

Input Voltage to the VSENSE terminal as Table 2-35.

4.

PRES ON, setting Power per Table 2-35 for the model being
calibrated.

5.

PRES OFF.

6.

LOAD ON, Adjust using the

…† keys until load DAM matches

Current Setting value.

MODEL

Power Setting

VSENSE Input

Current Setting

SLM-60-30-150 15.00W 50.00V 0.300A
SLM-60-60-300 30.00W 50.00V 0.600A
SLM-250-10-300 6.000W 30.00V 0.200A
SLM-500-10-300 30.00W 300.0V 0.100A
SLM-60-15-75 7.500W

50.00V 0.150A

SLH-60-120-600 60.00W 50.00V 1.2000A
SLH-60-120-1200 120.0W 50.00V 2.400A
SLH-60-120-1800 180.0W 50.00V 3.600A
SLH-60-240-1200 120.0W 50.00V 2.4000A
SLH-60-240-1800 180.0W 50.00V 3.6000A
SLH-60-360-1800 180.00W 50.00V 3.6000A
SLH-500-60-1800 180.00W 300.0V 0.6000A

Table 2-35. CP Mode, Level High, Offset Adjustment Settings (Item 35)

7.

LOAD OFF.

8.

Press the

© key to go to Item 36.

2-32

SLM/SLH Calibration Manual