Cp m – AMETEK SL Series Calibration Manual User Manual
Page 40
Calibration
Elgar Electronics Corporation
2.7.2
CP M
ODE
, L
EVEL
H
IGH
, O
FFSET AND
G
AIN
, I
TEMS
35
AND
36
1.
LOAD OFF
2.
Press the
© key to go to Item 35.
3.
Input Voltage to the VSENSE terminal as Table 2-35.
4.
PRES ON, setting Power per Table 2-35 for the model being
calibrated.
5.
PRES OFF.
6.
LOAD ON, Adjust using the
keys until load DAM matches
Current Setting value.
MODEL
Power Setting
VSENSE Input
Current Setting
SLM-60-30-150 15.00W 50.00V 0.300A
SLM-60-60-300 30.00W 50.00V 0.600A
SLM-250-10-300 6.000W 30.00V 0.200A
SLM-500-10-300 30.00W 300.0V 0.100A
SLM-60-15-75 7.500W
50.00V 0.150A
SLH-60-120-600 60.00W 50.00V 1.2000A
SLH-60-120-1200 120.0W 50.00V 2.400A
SLH-60-120-1800 180.0W 50.00V 3.600A
SLH-60-240-1200 120.0W 50.00V 2.4000A
SLH-60-240-1800 180.0W 50.00V 3.6000A
SLH-60-360-1800 180.00W 50.00V 3.6000A
SLH-500-60-1800 180.00W 300.0V 0.6000A
Table 2-35. CP Mode, Level High, Offset Adjustment Settings (Item 35)
7.
LOAD OFF.
8.
Press the
© key to go to Item 36.
2-32
SLM/SLH Calibration Manual