AMETEK SL Series Calibration Manual User Manual

Page 27

Advertising
background image

Elgar Electronics Corporation

Calibration

2.5.2

CR M

ODE

, L

EVEL

H

IGH

, R

ANGE

I O

FFSET AND

G

AIN

I

TEMS

23

AND

24

1.

LOAD OFF.

2.

Press the

© key to go to Item 23.

3.

Set the power supply voltage and the DC Voltage Standard to the
VSENSE Input value shown in Table 2-23 for the model being
calibrated.

4.

PRES ON.

5.

Set the current to the Current Setting value in Table 2-23 for the
model being calibrated.

6.

PRES OFF.

7.

LOAD ON.

8.

Adjust using the

…† keys until the DAM reading and the current

setting match.

MODEL Resistance

Setting

VSENSE Input Current Setting

SLM-60-30-150 3750

Ω 60.00V

0.016A

SLM-60-60-300 3750

Ω 60.00V

0.016A

SLM-250-10-300 15625

Ω 250.00V 0.08A

SLM-500-10-300 6250

Ω 500.0V 0.04A

SLM-60-15-75 5000

Ω 60.00V

0.012A

SLH-60-120-600 1875

Ω 60.00V

0.0320A

SLH-60-120-1200 1875

Ω 60.00V

0.0320A

SLH-60-120-1800 1875

Ω 60.00V

0.0320A

SLH-60-240-1200 937.50

Ω 60.00V

0.0640A

SLH-60-240-1800 937.50

Ω 60.00V

0.0640A

SLH-60-360-1800 625

Ω 60.00V

0.0960A

SLH-500-60-1800 15625.

Ω 500.00V 0.032A

Table 2-23. CR Mode, Level High, Range I Offset Adjustment Settings (Item 23)

9.

LOAD OFF.

10.

Press the

© key to go to Item 24.

SLM/SLH Calibration Manual

2-19