Cp m – AMETEK SL Series Calibration Manual User Manual
Page 38
Calibration
Elgar Electronics Corporation
2.7.1
CP M
ODE
, L
EVEL
L
OW
, O
FFSET AND
G
AIN
, I
TEMS
33
AND
34
1.
LOAD OFF.
2.
Press the
© key to go to Item 33.
3.
Input Voltage to the VSENSE terminal as Table 2-33 for the model
being calibrated.
4.
PRES ON, setting Power as table 38 for the model being calibrated.
5.
PRES OFF.
6.
LOAD ON, Adjust using the
Keys until the load DAM matches
Current Setting value.
MODEL
Setting Power
VSENSE Input
Current Setting
SLM-60-30-150 15.0W 5.00V 3.000A
SLM-60-60-300 30.0W 5.00V 6.000A
SLM-250-10-300 6.00W 30.00V 0.2A
SLM-500-10-300 30.00W 40.00V 0.1375A
SLM-60-15-75 7.50W
5.00V 1.500A
SLH-60-120-600 60.00W 5.000V 12.00A
SLH-60-120-1200 120.0W 10.000V 12.00A
SLH-60-120-1800 180.0W 15.00V 12.00A
SLH-60-240-1200 120.0W 5.00V 24.00A
SLH-60-240-1800 180.0W 7.500V 24.00A
SLH-60-360-1800 180.00W 5.00V
36.00A
SLH-500-60-1800 180.00W 300.00V 0.600A
Table 2-33. CP Mode, Level Low, Offset Adjustment Settings (Item 33)
7.
LOAD OFF.
8.
Press the
© key to go to Item 34.
2-30
SLM/SLH Calibration Manual