Calibration

Elgar Electronics Corporation

2.7.1

CP M

ODE

, L

EVEL

L

OW

, O

FFSET AND

G

AIN

, I

TEMS

33

AND

34

1.

LOAD OFF.

2.

Press the

© key to go to Item 33.

3.

Input Voltage to the VSENSE terminal as Table 2-33 for the model
being calibrated.

4.

PRES ON, setting Power as table 38 for the model being calibrated.

5.

PRES OFF.

6.

LOAD ON, Adjust using the

…† Keys until the load DAM matches

Current Setting value.

MODEL

Setting Power

VSENSE Input

Current Setting

SLM-60-30-150 15.0W 5.00V 3.000A
SLM-60-60-300 30.0W 5.00V 6.000A
SLM-250-10-300 6.00W 30.00V 0.2A
SLM-500-10-300 30.00W 40.00V 0.1375A
SLM-60-15-75 7.50W

5.00V 1.500A

SLH-60-120-600 60.00W 5.000V 12.00A
SLH-60-120-1200 120.0W 10.000V 12.00A
SLH-60-120-1800 180.0W 15.00V 12.00A
SLH-60-240-1200 120.0W 5.00V 24.00A
SLH-60-240-1800 180.0W 7.500V 24.00A
SLH-60-360-1800 180.00W 5.00V

36.00A

SLH-500-60-1800 180.00W 300.00V 0.600A

Table 2-33. CP Mode, Level Low, Offset Adjustment Settings (Item 33)

7.

LOAD OFF.

8.

Press the

© key to go to Item 34.

2-30

SLM/SLH Calibration Manual