Calibration

Elgar Electronics Corporation

11.

Set the power supply voltage and the DC Voltage Standard to the
VSENSE Input value shown in Table 2-26 for the model being
calibrated.

12.

PRES ON.

13.

Set the current to the Current Setting value in Table 2-26 for the
model being calibrated.

14.

PRES OFF.

15.

LOAD ON.

16.

Adjust using the

…† keys until the DAM reading and the current

setting match.

17.

Repeat steps 1 through 16 (press

ª to return to Item 25) until no

further adjustments are necessary for the DAM reading and the
current setting to match.

MODEL Resistance

Setting

VSENSE Input Current Setting

SLM-60-30-150

2.003Ω 60.00V

30.00A

SLM-60-60-300

1.001Ω 60.00V

60.00A

SLM-250-10-300

25.00Ω 250.00V

10.000A

SLM-500-10-300

50.00Ω 500.00V

10.000A

SLM-60-15-75

4.000Ω 60.00V

15.000A

0.5000 Ω 60.00V

120.00A

SLH-60-120-600
SLH-60-120-1200

0.5000 Ω 60.00V

120.00A

SLH-60-120-1800

0.5000 Ω 60.00V

120.00A

SLH-60-240-1200

0.2500 Ω 60.00V

240.00A

SLH-60-240-1800

0.2500 Ω 60.00V

240.00A

SLH-60-360-1800

0.167 Ω 60.00V

359.281A

SLH-500-60-1800

8.3333Ω 500.00V 60.00A

Table 2-26. CR Mode, Level Low, Range II Gain Adjustment Settings (Item 26)

2-22

SLM/SLH Calibration Manual