Example 5: P/F Test


P/F Operation hints you the signal is overstep its range according to the selected
High/Low Limit parameters; meanwhile the buzzer makes a sound to alarm (if
Sound is open)

Operating Steps:

1. Press

button to start AC Voltage Measurement and choose an appropriate

voltage range.

2. Connect test leads as Figure 2-8.
3. Set the High/Low Limit Value.
(1) Press

 P/F  High to set the high limit value.

Figure 3-9 Set the High Limit Value

(2) Press

 P/F  Low to set the low limit value.

Figure 3-10 Set the Low Limit Value

(3) Press ON to start P/F tesing. As shown in Figure 3-11, The test result is

“PASS” as the measured value is between 0V and 1V.

Figure 3-11 P/F Test result