RIGOL DM3058 User Manual

Page 29

Advertising
background image

RIGOL

User’s Guide for DM3058/DM3058E

2-7


3. Choose a proper Voltage range according to the measured circuit.


Table 2-1 Measurement Characteristics of DC Voltage

Ranges*

200mV, 2V, 20V, 200V, 1000V

Input Protection

1000V on all ranges (HI

terminal)

Configurable

Parameters

Range, DC input impedance, REL

NOTE

*

: All the ranges enable to obtain 20% value higher than original except 1000V.

Besides, both Manual and Auto are available for setting every range. When inputting range
is higher than 1000V at 1000V Level, “OVER LOAD” will be shown on the screen. 1000V
input protection exists in every range.

4. Set the DC input impedance

Press

Res to set the DC resistance as “10 MΩ (default value)” or “>10

GΩ”. Users can execute DC voltage measurement directly without modifying
this parameter which has been setup before leaving factory.

5. Set relative value (Optional)

Press REL to open or close Relative math function. When it is open, “REL” is

shown over the screen and the reading displayed is a relative value which
comes from the result of actual measurement value subtracts the value that
has been set. (For the details about how to set, please refer to

Math

Functions in Chapter 2.)


6. Read measurement result

Select required measurement (reading) rate by using Left or Right direction
keys and read the measurement result.

7. View history data

Press History to enter into the following interface and check the data getting
from this measurement and saved.

Figure 2-6 View interface of history data