User’s Guide for DM3058/DM3058E

2-38

Secondary Function

Secondary function button

is used to open double display, save current

instrument settings and open the REL setting interface quickly.

1. Open Double Display

Press

, this button will shine and the instrument is going to wait for the

secondary operation. By this time, if you press a function key, it will be
displayed in the Vice Display (for available combination, see table 2-13).

Table 2-13 Available Main/Vice Function Combination (shade is available)

Main Display Function

DCV DCI ACV ACI FREQ PERIOD 2WR 4WR Cap

Vice

Display

Function

DCV

DCI

ACV

ACI

FREQ

PERIOD

2WR

4WR

Cap

Explanations:

(1) If the same measurement function is used in both Main and Vice Display
 The readings in both of the display will update at the same time.
 If Math function (STA, REL, dBm and dB) is used in Main Display, when

opening Vice Display, the result will be shown in Main Display and the
measurement value before math operation is shown in Vice Display.

 If Math function (P/F) is used in Main Display, when open Vice Display, P/F

operation will be closed automatically and Vice Display will show the second
selected function normally.

(2) If different measurement functions are used in both Main and Vice Display
 The readings in both of the display will update alternately.
 If Math function (STA, P/F, dBm and dB) is used in Main Display, when open

Advertising