dB Measurement


The dB function applies to AC voltage and DC voltage measurements only.
Each dB measurement is different between the input signal and a stored relative
value, with both values converted to dBm.

Press

 dB, enter the menu shown below:




Figure 2-64 dB Setting

Table 2-19 dB Measurement Function Menu Function Description

Function Menu

Settings

Description

Default

The default value is 0 dBm.

ON

Open dB Measurement.

Status bar shows “dB”.

Save all changes, back to a higher level

menu.


The computation method of the dB:

dB
=10xLog

10

[(Reading

2

/ R

REF

) / 0.001W] – (dB setting value)


R

REF

expressed measuring the resistance value in the actual electric circuit.

Range of the dB setting value: -120 dBm ~ +120 dBm. The default is 0 dBm.

NOTE:


 Input a value in dB setting interface by direction button, and then store

it as dB setting value.

 Settings of dB value are stored in volatile memory.