Rosemount 5300 series – Emerson Process Management ROSEMOUNT 5300 User Manual

Page 105

Advertising
background image

Reference Manual

00809-0100-4530, Rev BA
July 2009

5-33

Rosemount 5300 Series

7. Select the TRANSDUCER1100 block and choose the Geometry tab.

8. Tank Height is the distance from the Upper Reference Point to the tank
bottom (see “Tank and Probe Geometry” on page 5-4). Make sure that this
number is as accurate as possible.

9. Select the Mounting Type used.

10. Select Inner Diameter if using pipe, chamber, or nozzle.

11. Type Nozzle Height if using nozzle.

F

OUNDATION

Fieldbus parameter:

TRANSDUCER 1100>GEOM_TANK_HEIGHT
TRANSDUCER 1100>MOUNTING_TYPE
TRANSDUCER 1100>PIPE_DIAMETER
TRANSDUCER 1100>NOZZLE_HEIGHT

12. Select the Environment tab.

13. Normally, the Measurement Mode does not need to be changed. The
transmitter is pre-configured according to the specified model.

Advertising