Rosemount 5300 Series

CHANGING A PROBE

Probe and Firmware
Compatibility

Transmitter heads with a firmware version earlier than 1.A4 (manufacturing
date before 2008-06-18) are not compatible with HP/HTHP/C probes marked
with R2.

Transmitter heads with a firmware version 1.A4 or later are compatible with
HP/HTHP probes without the R2 marking when the Trim Near Zone function
is performed, as illustrated below.

Only probes with the VC marking are compatible with the Dynamic Vapor
Compensation function. See Appendix C: Advanced Configuration to
determine if the Dynamic Vapor Compensation function is supported by the
transmitter head.

Table 7-1. Compatibility
between probe type and
firmware version

Compatibility Between Probe Types and Firmware Versions

Firmware Version

Probe Type

Standard

HP / HTHP without

R2 marking

HP / HTHP / C with

R2 marking only

(1)

HTHP with both

R2 and VC marking

(1)

Firmware version earlier
than 1.A4

Yes

Yes

No

No

Firmware version 1.A4

Yes

Yes

(2)

Yes

No

Firmware version 2.A2 or
later

Yes

Yes

(2)

Yes

(3)

Yes

(4)

(1) The R2 marking is on the housing seal or the adapter as seen in the figure.
(2) Trim Near Zone function is required.
(3) When Dynamic Vapor Compensation is not used.
(4) The probe requires that Dynamic Vapor Compensation is activated in the device for full functionality.

R 2

R 2

R 2

R 2

R2 VC

R2 VC

R2

R2

Advertising