5-25

Rosemount 5300 Series

Volume

8. Choose Volume Calculation Method

HART command: [2, 1, 5].

F

OUNDATION

Fieldbus parameters:

Calculation method:
TRANSDUCER 1300>VOL_VOLUME_CALC_METHOD

Tank Diameter:
TRANSDUCER 1300>VOL_IDEAL_DIAMETER

Tank Length:
TRANSDUCER 1300>VOL_IDEAL_LENGTH

Volume Offset:
TRANSDUCER 1300>VOL_VOLUME_OFFSET

To use volume calculation, choose a pre-defined calculation method based on
the tank shape that best corresponds to the actual tank. See “Volume
Configuration” on page 5-7.

Use the Strapping Table option if the actual tank does not match any of the
available options for pre-defined tanks, or if higher calculation accuracy is
desired.

Choose None if volume calculation is not desired at all.

The following standard tank shapes are available:

Vertical Cylinder

Horizontal Cylinder

Vertical Bullet

Horizontal Bullet

Sphere

None

The following parameters must be entered for a standard tank shape:

Tank diameter

Tank height/length (not for spherical tanks)

Volume Offset: use this parameter if you do not want zero volume and
zero level to match (for example if you want to include a volume below
the zero level)