Measuring range -12 interface -12, Rosemount 5300 series, Reference manual – Emerson Process Management ROSEMOUNT 5300 User Manual

Page 26: Measuring range, Interface

Advertising
background image

Reference Manual

00809-0100-4530, Rev BA

July 2009

Rosemount 5300 Series

2-12

Measuring Range

The measuring range differs depending on probe type and characteristics of
the application. Table 2-4 can be used as a guideline for clean liquids.

See Appendix A: Reference Data for the measuring range when using
Remote Housing.

Table 2-4. Measuring Range

The maximum measuring range differs based on application according to:

Disturbing objects close to the probe

Media with higher dielectric constant (

r

) has better reflection and a

longer measuring range

Surface foam and particles in the tank atmosphere might affect
measuring performance

Heavy coating / contamination on the probe may reduce the measuring
range and cause erroneous level readings

Disturbing EMC environment in tank

Tank material (e.g. concrete or plastic) for measurements with single
lead probes

Interface

Rosemount 5302 is the ideal choice for measuring the level of oil, and the
interface of oil and water, or other liquids with significant dielectric differences.
Rosemount 5301 can also be used for interface measurement in applications
where the probe is fully submerged in the liquid.

Figure 2-8. Interface
measurement with a Rosemount
5302 and a Rosemount 5301
(fully submerged probe).

Rigid Single Lead

Flexible Single Lead

(1)

Coaxial

Rigid Twin Lead

Flexible Twin Lead

Maximum Measuring Range

9 ft 10 in. (3 m) - for 8 mm probes
14 ft 9 in. (4.5 m) - for 13 mm probes

164 ft (50 m)

19 ft 8 in. (6 m)

9 ft 10 in. (3 m)

164 ft (50 m)

Minimum Dielectric Constant at Maximum Measuring Range

1.4 (1.25 if installed in a metallic
bypass or stilling well)

(1)(2)

1.4, up to 49 ft (15 m)

(1)

1.8, up to 82 ft (25 m)

(1)

2.0, up to 115 ft (35 m)

(1)

3, up to 138 ft (42 m)
4, up to 151 ft (46 m)
6, up to 164 ft (50 m)

1.2 (Standard)
1.4 (HP/C)
2.0 (HTHP)

1.4

1.4, up to 82 ft (25 m)

(1)

2.0, up to 115 ft (35 m)

(1)

2.5, up to 131 ft (40 m)

(1)

3.5, up to 148 ft (45 m)
6, up to 164 ft (50 m)

(1) The probe end projection software function will improve the minimum dielectric constant. Consult you local Emerson Process Management representative

for details.

(2) Measuring range may be lower depending on installation.

Level

Interface Level

Level = Interface Level

5302

5301

Advertising