Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual
Page 100
 
6F8A0917
- 99 -
Checking the preset counter setting
How to check the preset count value is shown below. 
To check the preset output function, select the function “I2: PRESET FNC.” 
Switch operation
Display example
Description
I1: PRESET CNT 
 00000300 
Select H1:PRESET from the setting item select menu. 
The currently set preset count value appears. 
I1: PRESET CNT
Returns to the menu of preset item selection.
Changing the preset count value
The following is an example to show how to change the preset value from 500 (count) to 1000 
(count) 
Switch operation
Display example
Description
I1: PRESET 
 00000500 
Select I1:PRESET from the setting item selection menu. 
The currently set preset count value (500 counts in this 
example) appears. 
I1: PRESET 
 00000500 
The cursor appears. 
Press
further to move the cursor to the desire digit to
change.
I1: PRESET 
 00001500 
 00001000 
Change the number of the digit. 
To change other digit, press
under this condition to move
the cursor to that digit and change the number there
 
I1: PRESET 
 00001000 
 
The display blinks and the set value is displayed for
confirmation
Press
MENU/ENT
again to write the data. Then the cursor
disappears and the display returns to the condition the setting
value is displayed.
I1: PRESET
Returns to the menu of setting item selection.