Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual
Page 62
 
6F8A0917
- 61 -
Changing the setting value for exciting frequency.
The following is an example to change the exciting frequency from 24Hz to 12Hz.
Switch operation
Display example
Description
A3: EX. FREQ. 
 24 Hz 
Select A3: EX FREQ. from the setting item selection menu. 
The currently set exciting frequency (24Hz in this example) 
appears. 
A3: EX. FREQ. 
 24 Hz 
The cursor appears.
A3: EX. FREQ. 
 12 Hz 
Change the exciting frequency to 12Hz. 
The exciting frequency appears cyclically between 6Hz, 12Hz 
and 24Hz. (Note 1) 
A3: EX. FREQ. 
 12 Hz 
The display blinks and the set value is displayed for
confirmation.
Press
MENU/ENT
again to write the data. Then the cursor
disappears and the display returns to the condition that the
setting value is displayed.
A3: EX. FREQ.
Returns to the menu of setting item selection.
Note 1: The exciting frequency is displayed cyclically as shown below.
6Hz
12Hz
24Hz
* Partially filled pipes type LF232*F : Useable only 6Hz