2 parameters check / change, 1 exciting current value – Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual
Page 57
 
6F8A0917
- 56 -
No.
Setting item
Display example
8.2.16
Preset count value 
Output function setting 
I1: PRESET CNT 
I2: PRESET FNC 
 00009000 
1:HOLD 
High/Low limit alarm On/Off 
High/Low limit alarm value 
J1: H ALM SET 
J2: H ALM VAL 
 1:ON 
+100.0 % 
8.2.17
High high/Low low limit alarm On/Off 
High high/Low low limit alarm value 
K1: HH ALM SET 
K2: HH ALM VAL 
 1:ON 
+110.0 % 
8.2.18
Fluid empty alarm
L1: EMPTY ALM
1:NORMAL
Self-diagnosis On/Off
L2: SELF CHECK
1:ON
8.2.19
Alarm output factor setting
L3: ALM PRESET
1:WITHOUT EMP
8.2.20
Rate-of-change limit value 
Control time 
M1: LIMIT RATE 
M2: LIMIT TIME 
 05.5 % 
 01 s 
8.4
Fixed value output (Loop check)
N1: FIXED OUT 
N2: FIXED CURR 
N3: FIXED PULSE 
 OFF 
 20.0 mA 
 1000 pps 
8.5.2
Zero offset adjustment
O1: MANUAL ZERO
-000.1 %
8.2 Parameters Check / Change
8.2.1 Exciting
Current
Value
The exciting current value is already adjusted when the product is shipped from the factory. Normally the 
exciting current value is set to the same value as described on the detector nameplate. 
Checking the exciting current value
Proceed as follows to check or change the exciting current value.
Switch operation
Display example
Description
A1: EX. CURR. 
 0.2100A 
Select A1: EX. CURR. from the setting item selection menu. 
The currently set exciting current value appears. 
A1: EX. CURR.
Returns to the menu of setting item selection.