6F8A0917
- 158 -
Check / Replacement of power supply unit
In general, electronic components deteriorate faster if the ambient temperature is high. The life of 
the power supply unit is about 10 years if the ambient temperature is 40°C, and 5 to 6 years if the 
temperature is 50°C or more. To extend the life of the flowmeter, it is recommended that you 
replace the power supply unit early. 
Contact your nearest Toshiba representative for a flowmeter inspection or unit replacement.
Product disposal
The main body or parts of the converter must be disposed of, according to the rules and regulations 
of your local government. 
Especially if you dispose of electrolytic capacitors to replace parts, have it done by an agency 
which is licensed to handle industry waste materials. 
Operative life
The operative life of this flowmeter is 10 years from the date of shipment.
The life of the flowmeter differs depending on the environmental conditions and the way it was 
used. To extend the life of the flowmeter, inspect the flowmeter periodically and clean or replace 
components if necessary.