Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual
Page 107
 
6F8A0917
- 106 -
Changing the fluid empty alarm On/Off setting.
The following is an example to change the fluid empty alarm setting from Alarm On (1:NORMAL) 
to Alarm Off (0:OFF). 
Switch operation
Display example
Description
L1: EMPTY ALM 
 1:NORMAL 
Select L1:EMPTY ALM from the setting item selection menu. 
Then the fluid empty alarm On/Off setting appears. 
L1: EMPTY ALM 
 1:NORMAL 
The cursor appears.
L1: EMPTY ALM 
 2:OFF 
Press
to select the fluid empty alarm Off
L1: EMPTY ALM 
 2:OFF 
The display blinks and the set value is displayed for
confirmation.
Press
MENU/ENT
again to write the data. Then the cursor
disappears and the display returns to the condition that the
setting value is displayed.
L1: EMPTY ALM
Returns to the menu of setting item selection.