12.2
Output When an Error or Alarm Occurs
Error indication
Measured value 
display 
Current output 
(4 to 20mA) 
Totalizer pulse 
output 
Remarks
ROM ERROR
(Note
1)
−
4mA (standard)
 (Note 
3)
Stopped
After power-up, measurement 
does not start. 
RAM ERROR
−
4mA (standard)
 (Note 
3)
Stopped
After power-up, measurement 
does not start. 
PARAMETER FAIL 
 (Note 
2)
Zero
4mA (standard)
 (Note 
3)
Stopped
−
EX.CURR OPEN
Zero
4mA (standard)
 (Note 
3)
Stopped
Still water zero adjustment 
cannot be conducted. 
EX.CURR ERROR
Zero
4mA (standard)
 (Note 
3)
Stopped
−
ADC ERROR
Zero
4mA (standard)
 (Note 
3)
Stopped
−
EMPTY
Zero
4mA (standard)
 (Note 
3)
Stopped
Still water zero adjustment 
cannot be performed. 
INVALID TOTAL
Measured value Measured value Measured value The error message disappears if
you clear (reset) the totalizer.
HIGH ALARM
Measured value Measured value Measured value
−
H-HIGH ALARM
Measured value Measured value Measured value
−
L ALARM
Measured value Measured value Measured value
−
L-LOW ALARM
Measured value Measured value Measured value
−
Note 1: The display and output may be uncertain depending on the nature of the ROM error. 
Note 2: If a parameter failure relating to the current output occurs, the current output may not become exactly 
the set value of the current output used when an alarm occurs.
Note 3: The output becomes the value selected in 8.2.11 “Current Output Setting Used When an Alarm
Occurs.”
Note 4: If the fluid empty alarm not included (WITHOUT EMP) is selected (standard) in the setting for alarm
output factors in 8.2.19 “Self-Diagnosis Function,” the output becomes 4mA and if the fluid alarm 
included (WITH EMP) is selected, the output becomes the same as the above in Note 3.