- 80 -
Changing the damping constant
The following is an example to show how to change the damping constant from 0.5 sec to 10 sec.
Switch operation
Display example
Description
D1: DAMPING 
 00.5 s 
Select D1: DAMPING from the setting item selection menu. 
The currently set damping constant (0.5 s in this example) 
appears. 
D1: DAMPING 
 00.5 s 
The cursor appears. 
Press
further to move the cursor to the desired digit to
change.
D1: DAMPING 
 10.5 s 
 
 10.0 s 
Change the number of the digit. 
To change other digit, press
under this condition to move
the cursor to that digit and change the number these.
D1: DAMPING 
 10.0 s 
The display blinks and the set value is displayed for
confirmation.
Press
MENU/ENT
again to write the data. Then the cursor
disappears and the display returns to the condition that the
setting value is displayed.
D1: DAMPING
Returns to the menu of setting item selection.
Note: If you set a value exceeding max value, it is forcibly changed to max value before data is
written.