7 remote still water zero adjustment – Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual
Page 147
                                    Advertising
                                    
                                        
                                    
                                
                                 
6F8A0917
- 146 -
10.7
Remote Still Water Zero Adjustment
Still water zero adjustment can be performed using an external signal.
To use this function, follow the procedure in 8.2.14, “Digital I/O Function” and set the DI as a zero 
adjustment start signal. 
[Signal input condition]
Zero adjustment signal
10 seconds
20 seconds
Zero
adjustment
start
H level signal
L level signal
* The signal must be set to H level first and then it must go to L level 10 seconds later but not more
than 20 seconds, and then zero adjustment starts at the falling edge of the signal. 
If the signal stays in H level less than 10 seconds or more than 20 seconds, the signal will be 
ignored. 