9.2 Converter Check / Calibration
9.2.1 0% Flow Rate Calibration
(Zero Calibration)
Using Mag-Prover internal calibration circuit, 0% flow rate (hereafter called zero point) calibration 
can be performed. 
Checking the zero point
Switch operation
Display example
Description
P1:FLOW CAL 0 
 0.0 % 
Select P1:FLOW CAL 0 from the calibration item selection 
menu. 
Zero point using the simulation input appears. 
P1:FLOW CAL 0
Returns to the menu of calibration item selection.
Calibrating the zero point
Switch operation
Display example
Description
P1:FLOW CAL 0 
 0.1 % 
Select P1:FLOW CAL 0 from the calibration item selection 
menu. 
Zero point using the simulation input appears. 
ADJUST READY 
 0.1 % 
Press
. Then the display shown left appears and the system
is ready for zero calibration. (Note 1)
P1:FLOW CAL 0% 
* CAL. 0% ADJ. 
The display shown left appears and zero calibration starts.
−
P1:FLOW CAL 0 
 0.0 % 
About 3 to 6 seconds later (Note 2), the zero calibration ends 
and a new zero point appears. 
P1:FLOW CAL 0%
Returns to the menu of calibration item selection.
Note 1: To cancel the adjustment from the condition of ADJUST READY, press
CANCEL
. The display returns
to the condition where zero point using the simulation input is displayed.
Note 2: Calibration time differs depending on the exciting frequency.