Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual
Page 112
 
6F8A0917
- 111 -
Changing the rate-of-change limit setting
The following is an example to change the limit value from 1.0% to 1.5%. 
To change the control time, select the function “M2: LIMIT TIME.” 
Switch operation
Display example
Description
M1: LIMIT RATE 
 1.0 % 
Select M1: LIMIT RATE from the setting item selection menu.
The currently set rate-of change limit value appears. 
M1: LIMIT RATE 
 1.0% 
The cursor appears. 
Press
further to move the cursor to the desired digit to
change.
M1: LIMIT RATE 
 1.5 % 
Change the number of the digit. 
To change other digit, press
under this condition to move
the cursor to that digit and change the number there.
M1: LIMIT RATE 
 1.5 % 
The display blinks and the set value is displayed for
confirmation.
Press
MENU/ENT
again to write the data. Then the cursor
disappears and the display returns to the condition that the
setting value is displayed.
M1: LIMIT RATE
Returns to the menu of setting item selection.
Note: If a value outside of the settable range is tried to set, *H. OVER SPEC. appears and the value returns to
the previously set value. Try to set a new value again.