- 111 -

Changing the rate-of-change limit setting

The following is an example to change the limit value from 1.0% to 1.5%.
To change the control time, select the function “M2: LIMIT TIME.”

Switch operation

Display example

Description

M1: LIMIT RATE
1.0 %

Select M1: LIMIT RATE from the setting item selection menu.
The currently set rate-of change limit value appears.

M1: LIMIT RATE
1.0%

The cursor appears.
Press

further to move the cursor to the desired digit to

change.

M1: LIMIT RATE
1.5 %

Change the number of the digit.
To change other digit, press

under this condition to move

the cursor to that digit and change the number there.

M1: LIMIT RATE
1.5 %

The display blinks and the set value is displayed for

confirmation.

Press

MENU/ENT

again to write the data. Then the cursor

disappears and the display returns to the condition that the

setting value is displayed.

M1: LIMIT RATE

Returns to the menu of setting item selection.

Note: If a value outside of the settable range is tried to set, *H. OVER SPEC. appears and the value returns to

the previously set value. Try to set a new value again.