Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual
Page 98
 
6F8A0917
- 97 -
Changing the pulse width
The following is an example to change the pulse width from 100ms to 250ms.
Switch operation
Display example
Description
H3: PLS WIDTH 
 100.0 ms 
Select H3: PLS WIDTH from the setting item selection menu. 
The currently set pulse width appears. 
H3: PLS WIDTH 
 100.0 ms 
The cursor appears. 
Press
further to move the cursor to the desired digit to
change.
H3: PLS WIDTH 
 200.0 ms 
 
↓
250.0 ms
Change the number of the digit. 
To change other digit, press
under this condition to move
the cursor to that digit and change the number there.
H3: PLS WIDTH 
 250.0 ms 
The display blinks and the set value is displayed for
confirmation.
Press
MENU/ENT
again to write the data. Then the cursor
disappears and the display returns to the condition that the set
value is displayed.
H3: PLS WIDTH
Returns to the menu of setting item selection.
Note: If you try to set a value exceeding the settable range, *H. OVER SPEC. appears and the value returns to
the previously set value. Try to set a new value again.