8.5 Zero Adjustment
8.5.1 Still Water Zero Adjustment
Still water zero adjustment must be performed when the fluid in the measuring pipe of the detector 
stays still. 
Switch operation
Display example
Description
−
 10.00 m/s 
 100.0 % 
Measured value being displayed (measurement mode)
ENTER111 
 000 
Pressing
MENU/ENT
in the measurement mode for 3 seconds
or more, the screen moves to the menu lock cancellation 
screen. 
* Pressing
CANCEL
in the mode selection screen, the mode
returns to the measurement mode.
ENTER111 
 111 
Set the menu lock cancellation cord [111] by 
pressing
and
.
* Pressing
CANCEL
in the mode selection screen, the mode
returns to the measurement mode.
>1:SET 
 2:CNT CTRL 
Press
MENU/ENT
for 3 seconds or more in the measurement
mode and then the mode changes to the mode selection screen.
A1:EX CURRENT
Select 1:SET in the mode selection screen and then the menu 
screen appears. 
E1: ZERO ADJUST
Press
and
so that the function number becomes E1:
ZERO ADJUST.
E1: ZERO ADJUST 
 01.0% 
The current flow rate measured value in % unit appears. 
(Check that the fluid to be measured stays still.) 
ADJUST READY 
 01.0% 
ADJUST READY appears and the system is ready for 
adjustment. (If “CANCEL” is pressed, Adjustment ready 
condition will be cancelled.) 
E1: ZERO ADJUST 
* ZERO ADJUST 
About 3 to 6 seconds later, the still water zero adjustment ends 
and the measured flow value after adjustment appears. 
−
E1: ZERO ADJUST 
 00.0% 
Returns to the menu of setting item selection.
E1: ZERO ADJUST
Press
and
so that the number digit of the function
number becomes 0.
Note2
Note2