Toshiba Tohsiba Electromagnetic Flowmeter Converter L5232 User Manual
Page 83
 
6F8A0917
- 82 -
Changing the low cutoff value
The following is an example to show how to change the low cutoff value from 1.0% to 3.0%.
Switch operation
Display example
Description
D2: LOW CUT 
 01.0 % 
Select D2: LOW CUT from the setting item selection menu. 
The currently set low cutoff value (1.0 % in this example) 
appears. 
D2: LOW CUT 
 01.0 % 
The cursor appears. 
Press
further to move the cursor to the desired digit to
change.
D2: LOW CUT 
 03.0% 
Change the number of the digit. 
To change another digit, press
under this condition to move
the cursor to that digit and change the number there.
D2: LOW CUT 
 03.0 % 
The display brinks and the set value is displayed for
confirmation.
Press
MENU/ENT
again to write the data. Then the cursor
disappears and the display returns to the condition that the
setting value is displayed.
D2: LOW CUT
Returns to the menu of setting item selection.
Note: If you try to set a value exceeding 10.0%, *H. OVER SPEC.* appears and the value returns to
the previously set value. In that case, try to set a new value again.