6F8A0917
- 38 -
Instructions
The operation principle of infrared switch is to irradiate infrared to the 
front of control panel and detect the reflection from finger when operating.
 
Normal operation is impeded depending on the conditions such as 
disturbing light from surroundings or stain attached to the control panel. 
When unable to avoid such condition, operate the EMF converter in the 
following manner. 
Remove the factor to impede proper operation of infrared switch as below:
・
Cover the control panel by hand so that light does not shine on it
・
Clean the stain attached on the control panel
・
Clean the stain on the finger or the gloves to operate the EMF 
converter, or wear gloves in light color 
・
When there is a reflecting object placed opposing to the control panel, 
stand in-between the reflecting object and the control panel to block 
the light 
Following are considered as the factors to impede infrared switch to 
operate properly. 
・
Intense light such as direct sunlight and reflected sunlight by window 
glass or metal plate 
・
Place where brightness changes always such as ON/OFF of lighting
・
Dense smoke or steam near the control panel
・
Those attached on the control panel such as rain (dew drop), snow, 
ice, mud and oil, and haze due to their attachment 
・
Operation of the control panel by hands wearing gloves in dark color 
or stained fingers and gloves 
・
Light reflecting object near the control panel, or reflecting object such 
as metal plate placed opposing to the control panel