44

MAIN MENU

CHANGE MODE

Enter

STATISTICS SETUP

STATISTICS SETUP

LIBRARY NAME

FUNCTION LINK

OFF

FILLING

CHECK WEIGHING

ANIMAL WEIGHING

RETURN TO SETUP

LIBRARY NAME

STATS1

STD DEV

ON

MEAN

ON

SUM

ON

MAXIMUM

ON

MINIMUM

ON

DIFFERENCE

ON

RELATIVE DEV

ON

Function links when set up to a particular function such as FILLING, CHECK
WEIGHING, or ANIMAL WEIGHING can be run directly from the Library
listing. Scrolling through the Library will reveal which item(s) are linked to
statistics.

From main menu

Enter CHANGE MODE and select STATISTICS, display advances to LIBRARY
NAME.

Library name

Enter a library name, up to 8 characters and press Enter button. Display
advances to STATISTICS SETUP with STD DEV highlighted. If a previous name
was entered and selected, the balance uses the parameters which were
previously set.

Turning statistics on or off

Select each option under the STATISTICS SETUP and turn either ON or OFF.
When you reach FUNCTION LINK, you have a choice of linking the statistical
data to either CHECK WEIGHING, FILLING or ANIMAL WEIGHING or not
linking and leave the FUNCTION LINK OFF.

Selecting a function link

Selecting FUNCTION LINK offers a choice of either FILLING, CHECK WEIGH-
ING, or ANIMAL WEIGHING functions to have statistics available on printouts.
Select the desired function and press Enter button.

Running the program

Select RUN and press Enter button. If statistics was selected to link with any
function, the display for that function will indicate STATS ON at the top of the
display.

Statistical result display

When a program has been run, a typical display as shown appears.

STATISTICS RESULT

NUMBER OF SAMPLES: 3

UNIT: GRAMS

MEAN: 203.571

MAXIMUM: 203.929

MINIMUM: 203.283

STD DEVIATION: 0.310

RELATIVE DEV: 0.064

SUM: 610.713

DIFFERENCE:0.546

STATISTICS SETUP

DIFFERENCE

ON

RELATIVE DEV

ON

AUTO SAMPLE

OFF

AUTO PRINT

OFF

SAMPLE SIZE

3

FUNCTION LINK

OFF

RUN

6.8 Statistics/Function link

Statistics are used when it is desired to compare a number of samples and examine the relative deviation of the
samples along with other statistical data. A minimum of three samples is required in this program. Statistics con-
tains menu options which include standard deviation, mean, sum, maximum, minimum, difference, relative devia-
tion, auto sample and sample size. Most of these can be set ON or OFF except sample size which can be set for a
particular number. When a printer or computer is connected to the balance, all statistical information can be ob-
served and printed. FUNCTION LINK offers a choice of three functions; CHECK WEIGHING, FILLING and ANIMAL
WEIGHING which can be linked directly to statistical data. Statistics can be enabled either through the CHANGE
MODE menu or the LIBRARY. The library method will automatically run the link. Changes can be made through the
CHANGE MODE menu setup.