4 user calibration, Procedure – Ohaus Voyager Balances User Manual

Page 16

Advertising
background image

10

Go Back

Help

Print

Enter

• Press

arrow button to select MAIN MENU.

• Press Enter button.

• Using the arrow buttons, select CALIBRATION.

• Press Enter button.

• Press

or

arrow button and select USER.

• Press Enter button, USER CALIBRATION is dis-

played.

• Using the arrow buttons, enter a value which is at

least 25% of the full span value, press Enter button.
The sample displays shown used a 2kg mass on an
4.1kg

balance.

• Clear the pan and press Enter button.

• Place the selected mass value on the pan and press

Enter button.

• Display indicates if calibration was successful and the

difference between the last calibration.

• Press any button, balance returns to weighing mode.

PROCEDURE

Go Back

Help

Print

Enter

MAIN MENU

CHANGE UNITS

SET BALANCE

RETURN TO

WEIGH

CUSTOM MENU

CHANGE MODE

CALIBRATION

MAIN MENU

CHANGE UNITS

SET BALANCE

LIBRARY

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN
LINEARITY

USER

CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

MAIN MENU

CHANGE UNITS

SET BALANCE

RETURN TO

WEIGH

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN
LINEARITY

USER

CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

ENTER CAL VALUE

UP/DOWN TO SCROLL CHARACTERS
LEFT/RIGHT TO MOVE CURSOR
ENTER TO ACCEPT
BACK TO QUIT

2000

MAIN MENU

CHANGE UNITS

SET BALANCE

RETURN TO

WEIGH

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN
LINEARITY

USER

CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

ENTER CAL VALUE

UP/DOWN TO SCROLL CHARACTERS
LEFT/RIGHT TO MOVE CURSOR
ENTER TO ACCEPT
BACK TO QUIT

0

USER CALIBRATION

PLEASE CLEAR THE PAN
AND PRESS ENTER KEY . . .

MAIN MENU

CHANGE UNITS

SET BALANCE

LIBRARY

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN
LINEARITY

USER

CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

ENTER CAL VALUE

UP/DOWN TO SCROLL CHARACTERS
LEFT/RIGHT TO MOVE CURSOR
ENTER TO ACCEPT
BACK TO QUIT

0

USER CALIBRATION

PLEASE PUT: 2000 G ON PAN.
AND PRESS < ENTER > . . .

MAIN MENU

CHANGE UNITS

SET BALANCE

RETURN TO

WEIGH

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN
LINEARITY

USER

CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

ENTER CAL VALUE

UP/DOWN TO SCROLL CHARACTERS
LEFT/RIGHT TO MOVE CURSOR
ENTER TO ACCEPT
BACK TO QUIT

0

USER CALIBRATION

PLEASE WAIT . . .

MAIN MENU

CHANGE UNITS

SET BALANCE

LIBRARY

CUSTOM MENU

CHANGE MODE

CALIBRATION

CALIBRATE

SPAN
LINEARITY

USER

CALIBRATION TEST
AUTOCAL
EXIT TO WEIGH

ENTER CAL VALUE

UP/DOWN TO SCROLL CHARACTERS
LEFT/RIGHT TO MOVE CURSOR
ENTER TO ACCEPT
BACK TO QUIT

0

USER CALIBRATION

USER CAL SUCCESSFULLY DONE!
THE LAST DIFFERENCE BETWEEN THIS CAL
AND LAST CAL IS:

0.01 GRAMS

PRESS ANY KEY TO EXIT . . .

3.3.4 User Calibration

User calibration is used when it is desired to calibrate the balance using a mass of known value. To use this calibra-
tion feature, proceed as follows:

0.00

WEIGHT

0%

100%

GRAMS

STABLE

MOVE HIGHLIGHT

MAIN MENU

CONTRAST

0.00

WEIGHT

0%

100%

GRAMS

STABLE

MOVE HIGHLIGHT

MAIN MENU

CONTRAST

Advertising