22

Calibration

4.3 User calibration

User calibration is used when it is desired to calibrate the balance using a mass of known value. To use this
calibration feature, proceed as follows:

CALIBRATE

SPAN

LINEARITY

USER

CALIBRATION TEST

AUTOCAL

AUTOCAL DELTA CORRECTION

EXIT TO WEIGH

From main menu

Scroll to CALIBRATION and press the Enter button. Display advances to
CALIBRATE with SPAN highlighted. Scroll to USER.

Enter

USER CALIBRATION

USER CALIBRATION RANGE:

100-400 GRAMS

PRESS <ENTER. TO CONTINUE.

User calibration range is specified.

Enter

USER CALIBRATION

PLEASE PUT 200.000 GRAMS

ON PAN AND PRESS <ENTER...

Enter

Place selected mass value on pan.

USER CALIBRATION

PLEASE WAIT. . .

THE DIFFERENCE BETWEEN THIS CAL

AND THE LAST CAL IS: 0.000 GRAMS

PRESS <ENTER> TO CONTINUE

Enter

Display indicates the difference between the present calibration and the
last calibration.

200.000

BASIC WEIGHING

Enter

0% 100%

Display indicates weight of the calibration masses. Remove masses from
pan. Balance returns to a weighing mode.

USER CALIBRATION

ENTER CAL VALUE IN GRAMS

200.000

200.000

Enter the calibration value.

GRAMS STABLE

PRESS <ENTER> FOR MENU

USER CALIBRATION

PLEASE CLEAR THE PAN. . .

AND PRESS <ENTER>.

Clear the pan.

Enter