Batch history result, Sample # 2, Remaining samples – Ohaus Voyager Balances User Manual

Page 294: Analyzing data, Viewing mean value trace, Symbol definitions, 53 balance modes, Sample value trace, Go back

Advertising
background image

53

Balance modes

Batch history result

The sample history result at the left represents a batch of three samples. To
run additional batches of samples, press the Enter button, the display will
return to PUT SAMPLE #1 ON PAN. Remember that only ten batches will be
displayed, however, if more batches are run, data is still recorded.

12.305

START/SAMPLE SETUP MAIN MENU

SQC - - SQC2

0%

100%

PUT SAMPLE #2 ON PAN

Sample # 2

Remove the first sample from the pan and place the second sample on the
pan. When the balance stabilizes, press the Enter button. When AUTO
SAMPLE is ON, you do not have to press Enter.

12.710

START/SAMPLE SETUP MAIN MENU

SQC - - SQC2

0%

100%

PUT SAMPLE #3 ON PAN

SAMPLES 3 BATCH NAME: NOI

UNIT GRAMS

NOMINAL WT: 12.181

MEAN 12.307

MAXIMUM 12.710

MINIMUM 11.906

STD DEVIATION: 0.402

RELATIVE DEVIATION: O.108

>+T2 0.00%

0

>+T1 0.00% 0

TI>N>-TI 100.00%, 3

<-T10.00%, 0

<-T20.00% 0

Remaining samples

Continue this procedure until all of your samples have been entered. When
all samples have been entered, a History screen is displayed which indi-
cates the parameters of the first batch.

SQC SETUP

AUTO SAMPLE

OFF

AUTO PRINT

OFF

NUMBER OF SAMPLES

3

VIEW MEAN VALUE TRACE

VIEW BATCH HISTORY

RUN

Analyzing data

Press the Go Back button until SQC SETUP is displayed. Select VIEW MEAN
VALUE TRACE. Press the Enter button and the Trace Value screen appears.
This screen indicates the limits you have set for your samples and the
amount of deviation per batch.

Viewing mean value trace

After you have taken the required number of data samples, you can view the
sample value trace and/or the sample batch history. Screen indicates 11
batches were processed with the last 10 displayed.

SQ1 TOTAL BATCHES I 1

05/24/99 08:06:28A 05/24/99 02:34:19
19.513 +T2
16.076 +T1
12.181 N
8.742 -Tl
7.500 -T2

Sample Value Trace

Symbol Definitions

The sample illustration at the left describes the symbols used on the Mean
Value Trace screen and their definitions.

Symbol definitions

MAXIMUM LIMIT

MINIMUM LIMIT

MEAN VALUE OF SQC SET

STANDARD

DEVIATION

Go Back

Advertising